Supply in Kazakhstan
KEG TRK supplies equipment diagnostics solutions (Ion Science) across Kazakhstan: Almaty, Astana, Karaganda, Ust-Kamenogorsk, Aktobe and other regions. Official warranty, monitoring system design, installation, commissioning and personnel training.
Key features
Applications and resources
Industry solutions, landing pages and articles on Ion Science MVI — portable mercury vapour detector
Typical tasks
- Surveying rooms and sites for mercury vapour
- Locating mercury spills
- Monitoring workplace exposure limits
- Recycling and fluorescent lamp reprocessing plants
- Mining
- Refining and hydrocarbon industries
- Bioremediation
Documentation
Official manufacturer PDF documents. Key specifications are also published on this page.
Need a price with delivery in Kazakhstan? We will prepare a quotation for your task.
Overview
The MVI (Mercury Vapour Indicator) is a compact portable mercury vapour indicator from the British company Ion Science. It shows mercury concentration in micrograms per cubic metre and responds to hazardous vapour in about three seconds. It is operated with one hand and suits surveys, screening for mercury spills and monitoring exposure limits.

The key advantage of the MVI is its dual beam ultraviolet absorption technology. It measures high mercury concentrations without saturating: no regeneration is needed between readings, so there is no downtime. After warm up it gives real time mercury vapour readings.
Ranges and alarm
- Two ranges — 0.1–199.9 and 1–1999 µg/m3, suited to time weighted average (TWA) monitoring.
- Audible alarm factory preset to 20 µg/m3. The level cannot be set above 35 µg/m3.
- 0–2 V output for a datalogger, which is not supplied.
Specifications
| Parameter | Value |
|---|---|
| Principle | dual beam UV absorption, 254 nm |
| Ranges | 0.1–199.9 and 1.0–1999 µg/m3 |
| Accuracy | ±5 µg or ±10% of reading |
| Repeatability | ±5% FSD |
| Zero drift | less than 5 µg/h |
| Response | approx. 3 s |
| Power | 15 V NiMH, more than 4.5 h from full charge |
| Conditions | +10 to +50 °C, 0–100% RH non-condensing |
| Dimensions | 145 × 295 × 80 mm (120 mm with handle) |
| Weight | 2.35 kg |
Interfering substances
The measurement is based on ultraviolet absorption, so other substances affect the reading. For example, at 100 ppm benzene gives a reading of 20 µg/m3, ethyl alcohol 6 µg/m3, toluene 3.5 µg/m3, and acetone and ethyl acetate 3 µg/m3 each. There is no measurable interference from carbon monoxide, carbon dioxide or ammonia.
Safety
Mercury is toxic if inhaled, ingested or absorbed through the skin or eyes, and amalgamates with gold, silver, stainless steel, aluminium and copper alloys. The instrument contains a 254 nm ultraviolet source; protective glasses are required if the lamp is operated exposed. The cover is to be removed by qualified technicians only.
Supply in Kazakhstan
KEG TRK is the official Ion Science representative in Kazakhstan. We supply mercury vapour indicators and arrange annual calibration and training. Price on request.
Technical specifications
| Principle | dual beam ultraviolet absorption, 254 nm |
| Ranges | 0.1–199.9 and 1.0–1999 µg/m3 (user selected) |
| Sensitivity | 0.1 µg/m3 and 1 µg/m3 |
| Accuracy | ±5 micrograms or ±10% of reading |
| Repeatability | ±5% FSD |
| Linearity | within 5% from 0 to 500 µg/m3 |
| Zero drift | less than 5 µg per hour |
| Response time | approx. 3 s |
| Alarm | audible, factory preset 20 µg/m3 (cannot be set above 35 µg/m3) |
| Power | 15 V NiMH rechargeable, more than 4.5 h from full charge |
| Output | 0–2 V for a datalogger (not supplied) |
| Operating conditions | +10 to +50 °C, 0–100% RH non-condensing |
| Dimensions | 145 × 295 × 80 mm (120 mm with handle) |
| Weight | 2.35 kg |
Applications
Manufacturer
Ion Science
Portable mercury vapour indicator from Ion Science: dual beam UV absorption, about 3 second response, ranges 0.1–199.9 and 1–1999 µg/m3, no saturation or regeneration between readings.
Key specifications
- Principle
- dual beam ultraviolet absorption, 254 nm
- Ranges
- 0.1–199.9 and 1.0–1999 µg/m3 (user selected)
- Sensitivity
- 0.1 µg/m3 and 1 µg/m3
- Accuracy
- ±5 micrograms or ±10% of reading
- Repeatability
- ±5% FSD
Official supplier in Kazakhstan
Related products

Ion Science Tiger XT — handheld PID detector for volatile organic compounds
Intrinsically safe Ion Science photoionisation detector for over 750 volatile organic compounds: 1 ppb to 20,000 ppm, T90 under 2 s, Fence Electrode technology against humidity and contamination, up to 24 h runtime and 120,000 logged readings.
- photoionisation detector (PID), Fence Electrode
- over 750 VOCs and toxic compounds

Ion Science Cub — personal VOC monitor
111 g personal PID monitor from Ion Science: volatile organic compounds up to 5000 ppm with resolution down to 0.001 ppm, a TAC version for benzene, STEL and TWA alarms, up to 12 h runtime, ATEX and IECEx.
- PID with Fence Electrode
- 10.6 eV; 10.0 eV TAC; 11.7 eV

Ion Science Falco 2 — fixed PID VOC detector
Flameproof fixed photoionisation detector for over 700 volatile organic compounds: diffused or pumped, ranges from 0–10 to 0–3000 ppm, a TAC version for benzene, 4–20 mA, RS485 Modbus and two relays.
- MiniPID 2 PID, Fence Electrode, Typhoon
- 10.6 eV; 10.0 eV (TAC — aromatics)
Need advice?