Supply in Kazakhstan
KEG TRK supplies non-destructive testing solutions (Olympus) across Kazakhstan: Almaty, Astana, Karaganda, Ust-Kamenogorsk, Aktobe and other regions. Official warranty, monitoring system design, installation, commissioning and personnel training.
Key features
Applications and resources
Industry solutions, landing pages and articles on Olympus 27MG — ultrasonic corrosion thickness gauge
Where it is used
Typical tasks
- Remaining wall thickness of corroded pipes, tanks and vessels
- Steel structures and tanks
- Maintenance rounds without data logging
- Incoming inspection of plate and pipe
Documentation
Official manufacturer PDF documents. Key specifications are also published on this page.
Need a price with delivery in Kazakhstan? We will prepare a quotation for your task.
Overview
The 27MG is a simple ultrasonic corrosion thickness gauge from Olympus (now the Evident brand). It measures remaining wall thickness from one side with a dual element transducer, so it suits pipes, tanks, vessels and steel structures.
It is built for fast work: the probe is recognised automatically, zero is compensated automatically, and Min/Max mode takes 20 readings per second to find the thinnest spot while scanning.

Modes
- Min/Max — 20 readings per second, shows minimum and maximum thickness.
- Differential — deviation from a set value.
- Gain adjust — for corroded surfaces.
- Freeze — holds the reading on screen.
Technical specifications
| Parameter | Value |
|---|---|
| Range | 0.50–635 mm |
| Resolution | 0.1 / 0.01 mm |
| Transducers | dual element 2.25–10 MHz |
| Display | backlit LCD |
| Power | 3 × AA, ~150 h |
| Protection | IP65 |
| Operating temperature | −10 to +50 °C |
| Dimensions | 84.0 × 152.4 × 39.6 mm |
| Weight | 340 g |
| Standard | EN15317 |
When you need a different instrument
For memory with PC transfer or measuring through paint, see the 45MG with options. For general-purpose gauging with B-scan, the 39DL PLUS.
What it is used for
The 27MG answers one question: how much metal is left in the wall. It is used for rounds of pipelines, vessels and tanks at oil and gas sites, power plants, metallurgy, chemical plants and concentrators, and for incoming inspection of plate and pipe. It is meant for staff who need a quick result without configuring complex modes.
How to use it in the field
- Clean the point of loose rust and scale and apply couplant.
- Connect the transducer. The gauge recognises it and applies the zero offset. The package includes a D7910 transducer, a test block and couplant.
- Check the reading on a sample of the same material and calibrate velocity if needed.
- Switch on Min/Max (20 readings per second) and sweep the transducer over the area, keeping the acoustic barrier across the pipe axis.
- Record the minimum value. The 27MG has no built-in memory, so readings go into a log.
Choosing between neighbouring models
| Parameter | 27MG | 45MG | 39DL PLUS |
|---|---|---|---|
| Range | 0.50–635 mm | 0.08–635 mm | 0.08–635 mm |
| Resolution | 0.1 / 0.01 mm | 0.1 / 0.01 / 0.001 mm | 0.1 / 0.01 / 0.001 mm |
| Memory | not listed | DL option | 792,832 readings |
| Through coating | not listed | EETC option | standard |
| Weight | 340 g | 431 g | 0.83 kg |
Choose the 27MG when you need a minimal, simple gauge for remaining thickness. When you need a log, coating measurement or single element transducers, move to the 45MG or 39DL PLUS.
What to prepare for an order
- the material and the range of wall thickness;
- surface temperature: the standard D7910 works at 0 to +50 °C, hot pipes need another transducer;
- pipe diameter, to pick the transducer size;
- whether extra transducers are needed for other ranges (2.25–10 MHz);
- calibration samples in your material.
Limits
- The manufacturer lists no data logger and no measurement through coatings.
- On a wall thinner than the transducer minimum the reading may be too high.
- It does not replace a flaw detector: it does not look for cracks or lack of fusion in welds.
Related materials
- How to Measure Pipe Wall Thickness with an Ultrasonic Thickness Gauge
- Ultrasonic Thickness Gauge: How It Works and Its Limitations
- Which Olympus thickness gauge to choose: 27MG, 38DL, 39DL, 45MG or 72DL
- Corrosion monitoring of pipes and vessels at power plants and refineries
- D790, couplant, step wedge
- Olympus thickness gauges and selection wizard, NDT equipment catalog, Thickness gauging and NDT: solutions by industry
Supply in Kazakhstan
KEG TRK supplies Olympus (Evident) thickness gauges and transducers and trains staff. Price on request.
Technical specifications
| Range | 0.50–635 mm (depends on material and conditions) |
| Resolution | 0.1 mm or 0.01 mm |
| Transducers | dual element 2.25–10 MHz, standard D7910 |
| Modes | Min/Max (20 readings/s), differential, gain adjust, freeze |
| Functions | automatic probe recognition, auto zero compensation |
| Display | large backlit LCD |
| Languages | nine, including Russian |
| Power | 3 × AA, about 150 h; 30 h continuous with backlight |
| Protection | IP65 |
| Operating temperature | −10 to +50 °C |
| Dimensions | 84.0 × 152.4 × 39.6 mm |
| Weight | 340 g |
| Standard | EN15317 |
Applications
Manufacturer
Olympus
Simple Olympus (Evident) ultrasonic corrosion gauge: 0.5–635 mm, 2.25–10 MHz dual element transducers, Min/Max mode at 20 readings per second, 150 h on three AA batteries, IP65, 340 g.
Key specifications
- Range
- 0.50–635 mm (depends on material and conditions)
- Resolution
- 0.1 mm or 0.01 mm
- Transducers
- dual element 2.25–10 MHz, standard D7910
- Modes
- Min/Max (20 readings/s), differential, gain adjust, freeze
- Functions
- automatic probe recognition, auto zero compensation
Official supplier in Kazakhstan
Related products

Olympus 45MG — ultrasonic thickness gauge
Olympus (Evident) ultrasonic thickness gauge: 0.08–635 mm, dual and single element transducers, Echo-to-Echo and THRU-COAT options for measuring through coatings, up to 475,000 readings, 3 AA batteries, IP67, 431 g.
- 0.08–635 mm (depends on material and transducer)
- 0.508–18.699 mm/µs

Olympus 39DL PLUS — ultrasonic thickness gauge
General-purpose Olympus (Evident) ultrasonic thickness gauge, successor to the 38DL PLUS: 0.08–635 mm, up to 60 Hz update, single and dual element transducers, THRU-COAT, Echo-to-Echo, B-scan, up to 792,832 readings, IP67, 0.83 kg.
- 0.08–635 mm
- 0.1 mm; 0.01 mm; 0.001 mm (option)

Olympus D790 — dual element transducer for thickness gauges
Olympus (Evident) dual element transducer for corrosion gauging: 5 MHz, 11 mm tip, 1.0–508 mm in steel, −20 to +500 °C, straight connector, potted cable.
- dual element
- 5 MHz

Olympus — step wedge for thickness gauge calibration
Step wedge with known step thicknesses for calibrating Olympus (Evident) ultrasonic thickness gauges and flaw detectors for velocity and zero. Material and step set are chosen to match the inspected part.
Need advice?
Our specialists will help you choose the optimal Olympus 27MG — ultrasonic corrosion thickness gauge configuration for your equipment.
Related articles
All articles →How to Measure Pipe Wall Thickness with an Ultrasonic Thickness Gauge
Corroded pipe wall thickness is measured with an ultrasonic gauge and a dual element transducer: clean the spot, apply couplant, calibrate sound velocity and zero, place the transducer with its barrier across the pipe axis and take the minimum reading.
Ultrasonic Thickness Gauge: How It Works and Its Limitations
How an ultrasonic thickness gauge calculates thickness from an echo, contact, delay line, immersion and dual element transducers, measurement modes, what affects accuracy and where the method fails.

Corrosion monitoring of pipes and vessels at power plants and refineries
How to build a field thickness-gauging route on TPP and refinery assets with Olympus 39DL/38DL

Which Olympus thickness gauge to choose: 27MG, 38DL, 39DL, 45MG or 72DL
Compare Olympus/Evident thickness gauges for corrosion, coatings and precision work — what fits power plants, refineries and mining sites