- Calibrating 27MG, 45MG, 39DL PLUS and 72DL PLUS gauges
- Accuracy checks in the field
- Thickness calibration of flaw detectors
Overview
A step wedge is a reference block with several areas of known thickness. It is used to calibrate an ultrasonic thickness gauge or flaw detector for zero and sound velocity and to check it before work. The wedge material and thickness set are chosen to match the inspected part.

Two-point calibration procedure
- Place the transducer on the thin step and perform zero calibration.
- Place the transducer on the thick step and calibrate velocity.
- Check the reading on an intermediate step.
Applications
- Calibrating 27MG, 45MG, 39DL PLUS and 72DL PLUS gauges
- Accuracy checks in the field
- Thickness calibration of flaw detectors
Related instruments
Gauges: 27MG, 45MG, 39DL PLUS. Transducers: D790. Couplant: ultrasonic couplant.
What it is used for
A step wedge is needed by anyone using an ultrasonic thickness gauge or flaw detector: it is used to set sound velocity and zero, and to check the instrument before work, after a transducer change or a change of material. Without calibration accuracy is no better than the velocity error. The standard package of the 27MG and 45MG already includes a two-step test block, and a separate one is selected for special materials and thickness ranges.
How to use it in practice
- Take a sample of the same material as the part, at the same temperature.
- Apply couplant, place the transducer on the thin step and calibrate zero.
- Place the transducer on the thick step, enter its thickness and calibrate velocity.
- Check the readings on an intermediate step and off the steps.
- Repeat calibration when the material, the transducer or the temperature changes noticeably.
Why the sample material matters
| Material | Sound velocity, m/s (Evident) |
|---|---|
| 1020 carbon steel | 5890 |
| 4340 alloy steel | 5850 |
| 302 stainless steel | 5740 |
| Aluminium | 6320 |
| Titanium | 6100 |
| Copper | 4660 |
The full table is in the article on sound velocity in materials.
What to prepare for an order
- the material of the part (steel grade or alloy);
- the range of thicknesses to measure: the steps must cover it;
- the type of instrument and transducer;
- conditions: temperature, where the sample will be stored;
- whether a separate instrument check under your procedure is needed.
Limits
- The sample must be the same material or one with a known sound velocity.
- Sound velocity depends on temperature, grain structure and alloy composition.
- A step wedge does not replace metrological verification of the instrument if your procedure requires it.
Related materials
- Sound Velocity in Materials: A Table for Thickness Gauging
- How to Measure Pipe Wall Thickness with an Ultrasonic Thickness Gauge
- Ultrasonic Thickness Gauge: How It Works and Its Limitations
- Thickness gauges: 27MG, 45MG, 39DL PLUS; transducer D790; couplant
- Olympus thickness gauges and selection wizard, NDT equipment catalog
Supply in Kazakhstan
KEG TRK supplies calibration blocks and Olympus (Evident) thickness gauges and selects a block for your material and thickness range. Price on request.
Applications
More by industry
Manufacturer
Olympus
Step wedge with known step thicknesses for calibrating Olympus (Evident) ultrasonic thickness gauges and flaw detectors for velocity and zero. Material and step set are chosen to match the inspected part.
Official supplier in Kazakhstan
Related products

Olympus 45MG — ultrasonic thickness gauge
Olympus (Evident) ultrasonic thickness gauge: 0.08–635 mm, dual and single element transducers, Echo-to-Echo and THRU-COAT options for measuring through coatings, up to 475,000 readings, 3 AA batteries, IP67, 431 g.
- 0.08–635 mm (depends on material and transducer)
- 0.508–18.699 mm/µs

Olympus 39DL PLUS — ultrasonic thickness gauge
General-purpose Olympus (Evident) ultrasonic thickness gauge, successor to the 38DL PLUS: 0.08–635 mm, up to 60 Hz update, single and dual element transducers, THRU-COAT, Echo-to-Echo, B-scan, up to 792,832 readings, IP67, 0.83 kg.
- 0.08–635 mm
- 0.1 mm; 0.01 mm; 0.001 mm (option)

Olympus 27MG — ultrasonic corrosion thickness gauge
Simple Olympus (Evident) ultrasonic corrosion gauge: 0.5–635 mm, 2.25–10 MHz dual element transducers, Min/Max mode at 20 readings per second, 150 h on three AA batteries, IP65, 340 g.
- 0.50–635 mm (depends on material and conditions)
- 0.1 mm or 0.01 mm

Olympus EPOCH 650 — ultrasonic flaw detector
Portable Olympus (Evident) ultrasonic flaw detector: tunable square-wave pulser up to 400 V, 0.2–26.5 MHz bandwidth, VGA display, DGS/AVG, DAC/TCG and AWS as standard, 100,000 IDs, 15–16 h runtime, IP67.
- 3.36–13,388 mm at 5900 m/s
- tunable square wave, 100/200/300/400 V
Need advice?
Our specialists will help you choose the optimal Olympus — step wedge for thickness gauge calibration configuration for your equipment.
Related articles
All articles →Sound Velocity in Materials: A Table for Thickness Gauging
Longitudinal ultrasonic velocity in steels, aluminum, copper, cast iron and plastics per Evident data (m/s and in/µs). Why velocity is calibrated on a sample and how temperature affects it.
How to Measure Pipe Wall Thickness with an Ultrasonic Thickness Gauge
Corroded pipe wall thickness is measured with an ultrasonic gauge and a dual element transducer: clean the spot, apply couplant, calibrate sound velocity and zero, place the transducer with its barrier across the pipe axis and take the minimum reading.
Ultrasonic Thickness Gauge: How It Works and Its Limitations
How an ultrasonic thickness gauge calculates thickness from an echo, contact, delay line, immersion and dual element transducers, measurement modes, what affects accuracy and where the method fails.